Sign in  
     
   
Search  
z z
 


UVV-501L Absorption Detector

a CEL-3100+UVV-501L-S.png
  Large view / Larger view
   

UVV-501L is a high performance UV/Vis absorption detector specially designed for Capillary Electrophoresis systems. It matches with CEL-3100 auto-sampler to form a complete system. UVV-501L uses the same core as UVV-500P which matches with the portable auto-sample CEP-5100 and CEP-5000 for field or lab analysis.

 

 

 

 

 

 

UVV-501 Specifications
Wavelength range 190-380nm (190-600nm optional)
Spectral half-width 8 nm
Accuracy of adjustment ± 1 nm
Reproducibility ± 0.5 nm
Light source Deuterium lamp, pre-aligned
Noise level at empty cell (250 nm, time constant 1 s) ± 1 x 10-5 AU
Drift at empty cell (for wavelength at 250 nm, constant room temperature, steady operation for 2 to 3 hrs) 5 x 10-4 AU/hr
Capillary O.D. 365 mm
Capillary I.D. 100, 75, 50, 25, 10 mm
Time constant 10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05, 0.04, 0.033, 0.02, 0.016, 0.01 sec.
Display 122x32 Graphic LCD
Output 1V/AU at scale = 1
Output scale 2, 1, 0.5, 0.2, 0.1, 0.05, 0.02, 0.01, 0.005, 0.002, 0.001
Interface RS232 (USB, TCP/IP optional)
Power supply 220VAC/50Hz, 115VAC/60Hz
Power input (wattage) 75 VA
Dimensions (W x H x D) 220(W) x 430(D) x 440(H) mm
Weight 9 kg

 

   
       
Copyright © EH Systems, Inc. All Rights Reserved