UVV-501 is a high performance UV/Vis absorption detector specially designed for Capillary Electrophoresis systems. It matches with CEL-3100 auto-sampler to form a complete system. UVV-501 uses the same core as UVV-500 which matches with the portable auto-sampler CEP-5100 for field or lab analysis.

The special CE flow cell has the same outer dimension as those for other detectors

With PKS Station control software and a RS232 cable (optionally USB or TCP/IP) it can also be used as a stand-alone detector.

UVV-501 Specifications
Wavelength range 190-380nm (190-600nm optional)
Spectral half-width 8 nm
Accuracy of adjustment ± 1 nm
Reproducibility ± 0.5 nm
Light source Deuterium lamp, pre-aligned
Noise level at empty cell (250 nm, time constant 1 s) ± 5 x 10-5 AU
Drift at empty cell (for wavelength at 250 nm, constant room temperature, steady operation for 2 to 3 hrs) 8 x 10-4 AU/hr
Capillary O.D. 365 µm
Capillary I.D. 100, 75, 50, 25, 10 µm
Time constant 10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05, 0.04, 0.033, 0.02, 0.016, 0.01 sec.
Display 122x32 Graphic LCD
Output 1V/AU at scale = 1
Output scale 2, 1, 0.5, 0.2, 0.1, 0.05, 0.02, 0.01, 0.005, 0.002, 0.001
Interface RS232 (USB, TCP/IP optional)
Power supply 220VAC/50Hz, 115VAC/60Hz
Power input (wattage) 75 VA
Dimensions (Width x Depth x Height) 224 x 415 x 436 mm
8.82 x 16.34 x 17.17 in.
Weight 7 kg

 

   
       