UVV-560 Absorption Detector Module

a UVV-550-S.png
  Large view / Larger view
   

UVV-560 is a high performance fixed wavelength UV/Vis absorption detector module specially designed for the dual-lifter Capillary Electrophoresis systems CEL-3000 and CEP-5000. Several standard UV light source configurations are used for this module. Refer to the specification table below.

The special CE Flow Cell has the same outer dimension as those for other detectors.

With PKS Station control software and a RS232 cable (optionally USB or TCP/IP) it can also be used as a stand-alone detector.

UVV-560 Specifications
Light source Mercury, Zinc or Cadmium lamps
or LED, UV-LED with adapters
Wavelength 254 nm with Hg
280,350,370 nm with coated Hg
214 nm with Zn
308 nm with Zn (other lines present)
229 nm with Cd
326 nm with Cd (other lines present)
UV-LED: 280, 365, 375 nm
350-1000 nm with Tungsten
Noise level at empty cell (254 nm Hg line, time constant 1 s) ± 2x10-5 AU
Drift at empty cell (for wavelength at 254 nm, constant room temperature, steady operation for 1 to 2 hrs) 4x10-5 AU/hr
Capillary O.D. 365 µm
Capillary I.D. 100, 75, 50, 25, 10 µm
Time constant 10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05, 0.04, 0.033, 0.02, 0.016, 0.01 sec.
Interface RS232 (USB, TCP/IP optional)
Power supply 12VDC/5A
Power input (wattage) 60 VA
Dimensions (Width x Depth x Height) 215 x 135 x 100 mm
8.46 x 5.31 x 3.95 in.
Weight 3 kg

 

   
       